Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Liquiline System CA80TN - Total nitrogen analyzer for precise online measurement according to HJ636

TN analyzer
Liquiline System CA80TN

Total nitrogen analyzer for environmental monitoring, industrial and municipal wastewater

CA80TN
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    UV photometric

  • Characteristic

    Process analyzer for total nitrogen in aqueous solutions
    UV photometric method following HJ636-2012

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39.2 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar

  • Sample flow rate

    Min. 15 ml/min (0.51 fl.oz/min)

  • Consistency of the sample

    Aqueous, homogeneous

  • Application

    Monitoring of wastewater treatment plant outlet
    Monitoring of the surface water quality
    Monitoring of industrial water treatment
    Monitoring and optimization of industrial wastewater treatment plants

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Modbus RS485/TCP (optional)
    Webserver (optional)
    Ethernet/IP
    Profibus DP
    Alarm relay

  • Input

    1 measuring channel
    1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0 to 10 mg/l N
    0 to 50 mg/l N
    0 to 200 mg/l N

  • Consumables

    Reagents and standard solutions CY80TN as well as cleaner CY800 are necessary for the operation
    Regular maintenance is done with the parts of the maintenance kit CAV880